9.9 Automatic Emissivity Setting
This function is used to make an automatic setting of emissivity using an external thermocouple connected with the
auxiliary input.
If necessary, enable the thermocouple input (see par. 9.2) and select the thermocouple type (see par. 8.6)
Press the <> or <> key until the thermocouple temperature measure is displayed
e
°C
Lock
Press <
ENTER
> key for setting the emissivity value with reference to the temperature values read from thermocouple
and pyrometer. The emissivity will change automatically to the correct value by the OSP.
e
°C
Lock
An error message appear if the
OSP
measure a temperature highest of the thermocouple. If the error happen, verify
the contact between the thermocouple and the object.
e
°C
Lock