-54-
10. OPERATIONS
10.2 Operations at Abnormal Inputs
1 Over-range input
The following table shows displays and printings for
input exceeding a printing range or a measuring
range.
• Measuring range:
Determined by the range No. programmed in
“Range/Printing range”. See Section 8.2.
• Printing range:
Trace printing range programmed in “Range/Printing range”.
Display Printing
区分
Input
Digital Bar graph Digital Analog
(1)
Lower than
minimum
measuring range*
_ _ _ _ _
□□□
……
□□□
(All go off.)
- - - - -
(2)
Lower than
minimum printing
range
Normal
□□□
……
□□□
(All go off.)
Normal
Overshooting to
minimum limit
(3)
Higher than
maximum printing
range
Normal
■■■
……
■■■
(All light.)
Normal
(4)
Higher than
maximum
measuring range*
_ _ _ _ _
■■■
……
■■■
(All light.)
+ - - - -
Overshooting to
maximum limit
* For the digital display and printing, the measured values can be displayed and printed up to about ±10% of the
span even when the values are outside the measuring range.
2 Input Disconnection
The display and printing, when the input is disconnected, differ depending on the “Burnout enable/disable”
programming. However, for channels of a voltage (mV or V) range, the burnout function is “disable (non)” even
when “Burnout Enable/Disable” is programmed. Program “Burnout enable/disable” to each channel. See
Section 11.11.
Display Printing
Programming burnout
Digital Bar graph Digital Analog
Disable
(non)
Undefined Undefined Undefined Undefined
Enable - downscale
(doWn bUrn)
□□□
……
□□□
(All go off.)
BURN
Overshooting to
minimum limit
Enable - upscale
(UP bUrn)
■■■
……
■■■
(All lights.)
BURN
Overshooting to
maximum limit
Definition of input
Printing range
Measuring range
(1) (2)
(3) (4)
Kommentare zu diesen Handbüchern